GE H
EALTHCARE
D
GA091568, R
IRECTION
8-4-10
Loading the Software - from DVD/CD
8-4-10-1
Introduction
The VIVID E9 software is delivered on:
•
one DVD for the VIVID E9 System Software
•
one CD for the VIVID E9 Application Software
•
extra CD(s) with software patch(es), may also be included.
A Software Patch CD is used when one or a few files should be replaced or added, but the change
doesn't require a complete software load. Example: Virus hardening software.
•
one CD with Printer Driver software
When installing the software, start with the System Software, then continue with the Application
Software and, if included, install the patch software after you loaded the Application Software.
The Printer Driver software must be loaded later, after the network connection has been verified or set
up.
8-4-10-2
Preparations for software loading
1.) If not already done, performe these tasks:
a.)
b.)
c.)
2.) Disconnect all external USB devices (USB Flash Card, USB Hard Driveetc.) before you start loading
the software.
8 - 16
5
EVISION
8-4-5 "Data Management - moving all images" on page 8-10
8-4-6 "Backing up the Patient Archive and System Configurations" on page 8-10
8-4-7 "Recording important settings and parameters" on page 8-11
Section 8-4 - Reloading the software
VIVID E9 S
M
ERVICE
ANUAL