Filter removal procedure
1.) Thread a long screwdriver through the extending cloth loop (tab).
2.) Pull down the screwdriver to release the filter from the unit.
Figure 8-201 Release of the bottom air filter
3.) Remove the filter.
DO NOT operate the unit without the air filters in place.
CAUTION
8-8-4-4
Filter Cleaning
You may either use a vacuum cleaner to vacuum the filter, or if needed, clean the filter by rinsing with
water.
Allow the air filter to dry thoroughly before re-installing it in the unit.
CAUTION
After cleaning the filter by rinsing with water, allow it to dry completely before re-installing.
When dry, install the filter as described next.
8-8-4-5
Filter installation procedure
Install the filter at the bottom of the VIVID E9. Align it in position (see: the tab in
magnets will help to position it correct.
Figure 8-202 VIVID E9 seen from below
5
EVISION
LINE UP THE FILTER SO THAT THE MAGNETS
ATTACH IN PLACE ON THE BOTTOM OF THE
VIVID E9.