Confirm The Level Of The Wash Or Rinse Bottle - Molecular Devices MultiWash+ User Manual

Microplate washer
Table of Contents

Advertisement

Confirm the Level of the Wash or Rinse Bottle

If the Liquid Warning option is enabled in the Setup menu, then the software displays a
prompt to make sure that the applicable wash or rinse bottle has sufficient liquid for the
process.
CAUTION! Running a prime or rinse process without sufficient liquid in the wash or
rinse bottle can damage the instrument. The minimum volume required for a
Prime/Rinse function is 7.5mL+1.5mL (total of 9mL), where 7.5mL is the volume of
liquid in the tube (from Rinse bottle to Dispense head of manifold) and 1.5mL is the
volume dispensed during the Prime/Rinse cycle. We suggest using a little more than
7.5mL + 1.5mL to have enough backup pressure to push liquid through the tube and
dispense the required volume (1.5mL).
To enable the Liquid Warning option in the Setup menu, see
Options on page
Figure 4-19: The Wash Liquid OK prompt
Table 4-12: Actions of Control Panel Buttons
Button
On the left
On the right
5032022 C
113.
Action
If the wash or rinse bottle does not contain enough liquid, press this button to return
to Prime/Rinse. See
Priming or Rinsing the Fluidics System on page
If the wash or rinse bottle contains enough liquid, press this button to go to the next
step. If no other options have been enabled, then you can define the volume of liquid
to use for the automatic prime or rinse process. See
the Prime or Rinse Process on page
The button has no function for this option.
The button has no function for this option.
The button has no function for this option.
Chapter 4: Priming or Rinsing the Fluidics System
Selecting Instrument Set Up
Defining the Liquid Volume for
47.
31.
45

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MultiWash+ and is the answer not in the manual?

Questions and answers

Table of Contents