SpectraMax ® GEMINI EM Dual-Scanning Microplate Spectrofluorometer Operator’s Manual 1311 Orleans Drive Sunnyvale, California 94089 Part # 0112-0088 Rev. A...
Page 2
2) All labor charges to repair the product for a period of one (1) year from the original date of delivery will be paid by Molecular Devices Corporation. 3) This warranty covers the SpectraMax GEMINI EM instrument only and does not extend to any computer, printer, software, reagents, or disposables used with the instrument.
Page 3
Molecular Devices Corporation service facility. You must call or write to the nearest Molecular Devices Corporation service facility to schedule warranty service. You may call Molecular Devices Corporation at the telephone number or address below to locate the nearest service facility.
Conventions The names of keys that appear on the SpectraMax GEMINI EM control panel are shown in boxed Helvetica type. Example: Used in this [Setup] Italic and boldface type are used for emphasis. Examples: “Press carefully to Manual engage,” “ Do not press down. ”...
Microplates having 6, 12, 24, 48, 96, and 384 wells can be used in the SpectraMax GEMINI EM. Top or bottom detection is available at the touch of a button. Detection of species collected on membrane plates is also possible.
Chapter 1 Component The main components of the SpectraMax GEMINI EM described in this Description manual are: • The control panel • The microplate drawer • Optical system • The back panel (connections and power switch) Back Panel Control Panel...
Page 15
45 seconds. If you plan to open the drawer during a kinetic read, first determine the minimum read interval allowed and then increase the setting by a minimum of 45 seconds. The drawer will close auto- matically after this interval before the next read. SpectraMax GEMINI EM Operator’s Manual...
(see Chapter 5, “Troubleshooting”). Microplates The SpectraMax GEMINI EM can accommodate standard 6-, 12-, 24-, 48-, 96-, and 384-well microplates. Black-walled, clear-bottom or all black microplates are generally recommended for fluorescence assays because they have lower backgrounds than clear plates.
Figure 1.4: Components of the SpectraMax GEMINI EM Optical System 1) The excitation light source is a xenon flashlamp. (Note that the lamp is off when luminescence mode is selected.) 2) The light passes through a band-pass filter which reduces the amount of stray light to the excitation monochromator.
Parallel Port Serial Port Label Figure 1.5: Components on the Back Panel of the SpectraMax GEMINI EM The following components are located on the back panel of the SpectraMax GEMINI EM: • Power switch : a rocker switch, labeled I/O (for on and off, respectively).
(gray area) and on to the PMT, resulting in a purer emission signal and more accurate data. The SpectraMax GEMINI EM allows scanning of both excitation and emission wavelengths, using separate tunable monochromators. One benefit of being able to scan emission spectra is that you can assess more accurately whether the emission is, in fact, the expected fluorophore, or multiple fluorophores,...
Functional The SpectraMax GEMINI EM is designed to be operated using SoftMax Pro Description software running on a computer connected to the instrument. Stand-alone functions are limited to setting and enabling temperature control and opening or closing the microplate drawer.
The SpectraMax GEMINI EM has been designed to regulate the temperature of the microplate chamber from 4°C above ambient to 45°C. Upon power up, when the incubator is off, the temperature in the SpectraMax GEMINI EM microplate chamber is ambient and isothermal. Turning on the incubator by...
Computer Control The SpectraMax GEMINI EM is equipped with an 8-pin DIN RS-232 serial port through which the computer communicates with and controls the instru- ment. (Different types of cables are available for connecting to different types of computers—see Appendix A, “Cables and Accessories.”) SoftMax Pro soft-...
Average lamp lifetime 2 years normal operation (estimate) Detector Photomultiplier (R-3896) Read time 96 wells in <15 seconds (measurement type may extend read time) Shaker Time 0–999 seconds Temperature control (chamber) Ambient +4°C to 45°C Ramp up to 37°C <20 minutes 1-14 SpectraMax GEMINI EM Operator’s Manual...
Page 25
Size (h × w × d) 13.5" (340 mm) × 16.5" (420 mm) × 16.5" (420 mm) Weight 35 lb (16 kg) Power consumption 500 VA maximum Line voltage and frequency 90–240 VAC, 50/60 Hz SpectraMax GEMINI EM Operator’s Manual 1-15...
A material which absorbs light energy of a characteristic wavelength, under- goes an electronic state change, and emits light of a longer wavelength. Gain The amount of increase in signal power expressed as the ratio of output to input. 1-16 SpectraMax GEMINI EM Operator’s Manual...
Page 27
SoftMax Pro An integrated software program (from Molecular Devices Corporation) that is used to control and collect data from the SpectraMax GEMINI EM instru- ment. Stokes Shift The difference between the wavelengths of the excitation and emission peaks.
Page 28
Chapter 1 1-18 SpectraMax GEMINI EM Operator’s Manual...
After examining the carton, place it on a flat surface in the upright position. Open the top of the box and lift the SpectraMax GEMINI EM, along with the packing materials around the ends, up and out of the shipping box. Remove the packing material from both ends of the instrument and set the instrument down carefully.
Chapter 2 3) Insert the female end of the power cord into the power receptacle at the rear of the SpectraMax GEMINI EM. Connect the male end to a grounded power outlet of the appropriate voltage. Molecular Devices recommends that you use a surge protector between the power cord and the grounded power outlet.
Adapter Incorrect insertion or removal of the adapter may cause dam- CAUTION: age to the microplate drawer of the SpectraMax GEMINI EM. 1) Turn power to the instrument on. 2) Press the button on the front panel or activate the drawer open [Drawer] command in SoftMax Pro software.
Page 34
Chapter 2 SpectraMax GEMINI EM Operator’s Manual...
Page 35
Chapter 3: Operation Prepare for a Reading....3-3 Turn the Instrument and Computer On 3-3 Set the Temperature ....3-3 Read the Microplate .
Page 36
Chapter 3 SpectraMax GEMINI EM Operator’s Manual...
Turn the Instrument and Computer On Reading The power switch for the SpectraMax GEMINI EM is located on the back panel. Press the rocker switch to the on position. The instrument will auto- matically perform diagnostic checks to ensure that it is functioning correctly.
Overview dures required to perform an assay using the SpectraMax GEMINI EM. 1) Turn on the power switch of the SpectraMax GEMINI EM (located on the back panel). The microplate drawer will open automatically. 2) If you wish to regulate the temperature inside the microplate chamber,...
Page 39
PMT voltage for varying concentrations of sample in the plate. • The chamber of the SpectraMax GEMINI EM is isothermal at ambient as well as at elevated temperatures. The temperature in the reading chamber may be adjusted from 4°C above ambient to 45°C.
570, 590, 610, 630, 665, or 695 nm. The cutoff value should be near the maximum emission wavelength (preferably between the excitation wave- length and the maximal emission wavelength) but at least 35 nm greater than the excitation wavelength. SpectraMax GEMINI EM Operator’s Manual...
Page 41
fixed at 485 nm and emission was scanned from 490 nm to 560 nm (buffer blanks are not shown in this plot). Table 3.1 following lists default settings for the emission cutoff filters. SpectraMax GEMINI EM Operator’s Manual...
Chapter 3 Figure 3.2: Effects of Cutoff Filters on Fluorescein. Emission was scanned from 490 to 560 nm; excitation was fixed at 485 nm. SpectraMax GEMINI EM Operator’s Manual...
Page 43
530 - 549 550 - 569 570 - 589 590 - 609 610 - 629 630 - 664 665 - 694 695 - 850 Note: For spectrum mode, the default is "manual" (no automatic cutoff). SpectraMax GEMINI EM Operator’s Manual...
Page 44
Chapter 3 3-10 SpectraMax GEMINI EM Operator’s Manual...
If you have any problems using the SpectraMax GEMINI EM Dual-Scanning Microplate Spectrophotometer, in the U.S., contact our Technical Services group at 1-800-635-5577; elsewhere contact your local representative.
SpectraMax GEMINI EM. If power failed while the SpectraMax GEMINI EM was already on, check that the power cord is not loose or disconnected and that power to the power outlet is functioning properly.
7) It is possible that only one of the fuses may have blown. Molecular Devices recommends that you replace both fuses, however, to ensure con- tinued proper operation. Pull both fuses out of the holder and discard them. SpectraMax GEMINI EM Operator’s Manual...
Adapter Incorrect insertion or removal of the adapter may cause CAUTION: damage to the microplate drawer of the SpectraMax GEMINI EM. The corner cut-out must be in the lower left corner where the plate pusher is located. 1) Turn power to the instrument on.
SoftMax Pro software. Figure 4.5: Microplate Drawer without Adapter 3) Remove the adapter plate. Moving the If you need to relocate the SpectraMax GEMINI EM, follow these steps. SpectraMax GEMINI EM The SpectraMax GEMINI EM weighs approximately WARNING: 35 pounds (16 kilograms).
Page 52
Chapter 4 SpectraMax GEMINI EM Operator’s Manual...
This chapter lists error codes that may occur while using the instrument, fol- lowed by their most likely causes and remedies. Maintenance procedures are described in Chapter 4. For problems with the SpectraMax GEMINI EM that are not listed here, in the U.S., contact Molecular Devices Technical Services group at 1-800-635-5577;...
Chapter 5 For all other error messages (codes not listed in the table above), please con- tact your local Molecular Devices representative for assistance. Opening the If an error occurs while the drawer is closed and you need to remove a micro-...
Cables and Accessories Cables Molecular Devices recommends that you use high-quality, double-shielded cables to connect the SpectraMax GEMINI EM to the computer. Choose cables that meet the following requirements: Serial Interface Cable (contact Molecular Devices for specific pin-out requirements) Macintosh: Male DB8 to male DB8...
Page 60
Appendix A SpectraMax GEMINI EM Operator’s Manual...
Page 61
Appendix B Common Wavelengths for Fluorescence and Luminescence Fluorescence ......B-3 Luminescence ......B-3...
Page 62
Appendix B SpectraMax GEMINI EM Operator’s Manual...
Page 63
Common Wavelengths for Fluorescence and Lumines- Values in this table are based on the literature. You may wish to scan NOTE: your fluorochrome of interest in the SpectraMax GEMINI EM to determine the optimal excitation and emission wavelengths for your application.
Page 64
Appendix B SpectraMax GEMINI EM Operator’s Manual...