7. Carefully lift the instrument out of the box.
8. Remove the foam and cardboard packing materials from the instrument.
Figure 1-5: Instrument Packing Materials
9. Remove the plastic wrapper from the instrument.
10. Gently place the instrument on a dry, flat, vibration-free area of the workbench.
The instrument is 21.00 cm (8.27 in.) wide, 21.00 cm (8.27 in.) high, and 46.00 cm
(18.11 in.) deep. To provide access to disconnect power from the instrument, maintain a
20 cm to 30 cm (7.9 in. to 11.8 in.) gap between the rear of the instrument and the wall.
Keep the instrument away from direct sunlight.
11. Visually inspect the instrument for loose, bent, or broken parts. If damage is found,
immediately report it to your Molecular Devices representative.
12. Compare the serial number on the rear of the instrument with the instrument serial
number on the shipping document and the packing list.
13. Make sure that all the instrument accessories are present and match the shipping
document and the packing list.