Making the connection
Positioning the access point
63 98 650 D 3439
D 3439.201.07.10.23 10.2017
5.4.7.5

Connecting to the PC via WLAN (option)

Connect access point
➢ Connect the LAN port A of the milling and grinding unit and the
access point, using the network cable (10m, Order No.: 61 51 521).
 The access point is pre-configured at the factory for this
application.
1. As a test, place the access point near the milling and grinding unit at
head level or higher.
2. Perform a communication test as described in the separate
instructions (see "Operating MC XL via WLAN in infrastructure
mode", chapter "Final work, analyzing connection quality"). If
applicable, follow the instructions on changing channels.
3. After you have found the optimum setting, take the acquisition unit
and place it in the position in which it will be operated that is farthest
away from the access point.
4. From this position, repeat the communication test you conducted
earlier. If the results are satisfactory, leave the access point
permanently in this position.
5. If the results are not satisfactory, position the access point outside of
the room in which the milling and grinding unit is located and repeat
the communication test.
 If the connection quality is still not adequate, WLAN
communication cannot be easily achieved under the local
conditions. In this case, ask your network administrator for
assistance.
NOTICE
LAN connection
Operation via a cable LAN connection is possible at any time.
5 Installation and startup