Operating Instructions CEREC MC XL
61 46 885 D 3439
D 3439.201.02.11.02 10.2009
Make sure that the ventilation slots underneath and at the back of the unit
remain unobstructed. The distance between the rear side of the unit and the
room wall must be at least 10 cm.
Note that the unit weighs 43 kg!
The unit must not be installed at sites with a high level of humidity or dust!
CAUTION
Installation in a cabinet
If the unit is installed in a cabinet, you must provide for adequate heat
exchange.
The ambient temperature surrounding the unit must be between 5°C and
40°C.
5.3.1

Only with integrated scanner (option)

CAUTION
Impairment of the scanned result
Impairment of the scanned result due to sudden incidence of light.
A sudden, strong incidence of light may falsify the scanned result.
Set the unit up so that the milling chamber is not located directly in the
beam path of an extreme light source and is not exposed to direct sunlight.
5.4

Initial startup

CAUTION
Important information on initial startup
Observe the software installation instructions!
5 Installation and startup
Initial startup
13