3
Type a file name in the
• Number of characters of a file name: Up to 100 characters
• Maximum length of a file name including its path: 255 characters
4
Set [Waveform].
Off
Does not save waveforms.
On
Saves waveforms.
[Type]
Waveform
Saves waveform data in binary format.
(Binary)
(Only data saved in binary format can be loaded onto the instrument)
Waveform (Text)
Saves waveform data in text format.
(The instrument decimates the data and saves the remaining data as a file. The files
can be opened with editors and spreadsheet software installed in PCs; however, they
cannot be loaded onto the instrument.)
[Method]
Set the countermeasure method to be taken if the size of waveform data to be saved exceeds the capacity of
the storage device.
Normal
Stops the auto-save if the storage device is full.
Delete
Deletes old files and continues the auto-save even when the storage device is full
(waveform files only).
The instrument deletes files created after the measurement has started; it does not
previously existing files.
[Channel]
All
Saves the data acquired through all measured channels (channels with measurement
set to [On]).
Saves the data acquired through the channels the waveform display of which is set to
[Off].
Display
Saves the data acquired through the channels the waveform display of which is set to
[On]
(when
[Divide]
[Type]
Off
Saves a file without dividing it.
16
MB,
Select this option to divide a large file into several files and save them. A file is divided
32
MB,
into several files each of which has the specified size. The instrument creates a folder
64 MB
with the specified name, divides a file into several files, and then saves them in the
folder.
A folder is automatically created, and then waveform files and an index file (extension: .IDX) are created in the
folder. Loading the IDX file allows the waveform files to be loaded in a batch.
[Divide]
[Type]
(when
Off
60,000
data,
1,000,000 data
[File name]
box.
on all sheets.
is set to
[Waveform (Binary)]
[Waveform
(Text)])
is set to
Saves a file without dividing it.
Divides a file into several files each of which contains the specified number of data
points.
)