8. Either set the RUN 1 time or press the ENTER soft key (see "Run Timer" on page 5-2).
Pressing the ENTER soft key at the final timer value automatically returns you to the
output display screen.
Note: Press the MODIFY knob at any time to return to the output display screen; note that
this disables the recipe (sets the number of steps to zero).

PROGRAMMING OUTPUT SETPOINT LEVEL

See "Output Setpoint Level" on page 5-3 for a description of this function.
TIMERS
RAMP/RUN
To program one or more ramp/run steps, follow these steps:
1. At the top-level menu, press the TIMERS soft key.
2. Press the RAMP/RUN soft key.
3. Set the number of steps (up to five) by turning the MODIFY knob.
a. If you select only one step, or if the RUN 1 timer is set to zero, press the ENTER soft
key to return to the output display screen automatically.
b. If you select more than one step, and the RUN 1 timer is not set to zero, press the
ENTER soft key to cause RAMP 2 TIMER and so on to be displayed until you have set
all values or scrolled past them with the ENTER soft key.
4. Either set the RAMP 1 timer or press the ENTER soft key. You can also change the
method of regulation at this point.
5. Set the SETPOINT 1 value by turning the MODIFY knob; select a method of regulation
by pressing the appropriate REGULATION switch.
6. Press the ENTER soft key.
7. Either set the RUN 1 timer or press the ENTER soft key.
Pressing the ENTER soft key at the final timer value automatically returns you to the
output display screen.
Note: Press the MODIFY knob at any time to return to the output display screen; note that
this disables the recipe (sets the number of steps to zero).
To specify a manual setpoint, follow these steps:
1. Disable the recipe by setting the number of steps to zero.
If the recipe is not disabled, the message "Timers Active" displays. If the recipe is
disabled, "Setpoint = nn.nn" displays.
2. Turn the MODIFY knob until you reach the desired setpoint.
When you turn on output, the MDXII unit ramps to that setpoint in 600 ms (50 ms on
units with fast ramp/run).
5700262-P
Number of steps
Panel Connection, Controls and Indicators
MDXII Power Supplies
setpoint n =
6-7