An "N" on the label plate on the rear of the unit indicates that the neutral is internally
connected. You must connect your neutral wire to this N terminal. If the label plate shows no
"N," no neutral connection is required. A specific phase rotation is not required on any model.
The input connector on the MDXII 30k unit is a five-position, high-current, pluggable
connector with the same line labels as are on the 15k unit.
Note: For both the MDXII 15k unit and 30k unit, connect input power to terminal block
contacts with stranded wire of a gauge recommended by local building codes.
To connect to the power supply's input power connector, follow these steps:
DANGER:
Before making any other connection, connect the protective earth terminal to a
protective earth-ground.
1. Strip approximately 60 to 70 mm of the insulation from the end of the power cord.
2. Separate the five wires and strip approximately 10 mm of the insulation from each wire.
3. Ensure that the yellow and green ground wire is approximately 7 mm longer than the other
wires.
4. Remove the cover from the strain relief.
5. Slip the cable through the cord grip.
6. Connect each wire to the correct input terminal block according to the labels.
7. Tighten the cable clamp on the line cord.
8. Slide the drawer back in place and tighten the two drawer screws.
4-8
Installation
5700262-R