Serial Port Indicators; Ae Bus Indicators; Command Status Response (Csr) Codes; Table 7-10. Ae Bus Command-Status Response (Csr) Codes - Advanced Energy MDX II User Manual

Table of Contents

Advertisement

SERIAL PORT INDICATORS

AE Bus Indicators

COMMAND STATUS RESPONSE (CSR) CODES

When you invoke a command through a serial port using an AE Bus protocol, a value is
returned in response to the command. The meaning of each value is explained in Table 7-10 on
page 7-57.

Table 7-10. AE Bus Command-Status Response (CSR) Codes

Value
Meaning
0
Command accepted
1
Command not accepted because the MDXII unit is not in valid control
mode
2
Command not accepted because output is on
3
Command not accepted because output is off
4
Command not accepted because the requested value is outside the
allowable range
5
Command not accepted because User pin 14 (RMT OFF.D) is high (page
9-11)
6
Command not accepted because the program source for this parameter is
external (page 5-16)
7
Command not accepted because a fault condition has not been
acknowledged
8
Command not accepted because the MDXII unit is ramping
9
Command not accepted because the data byte count is incorrect
10
Command not accepted because the User port has priority
11
Command not accepted because more than 50 V is present at the MDXII
unit output
12
Command not accepted because this feature is not available on your unit
13
Command not accepted because the control panel has priority
14
Invalid regulation mode
15
Invalid control mode (same as Value 1 above)
16
Target life consumed
17
Recipe is running
18
Ramping is underway
5700262-R
Serial Port Connection, Controls and Indicators
MDXII Power Supplies
7-57