®
Advanced Energy
Using More Functions
SPECIFYING A TAP SETTING
See "Selecting the Correct Tap" on page 5-8 for a description of this function.
⇒
MORE
TAPSET
Note: If you try to change taps when more than 50 V is present at the MDXII unit's output, the
message "Latent Output Voltage Exists" appears briefly and then clears. You are not
allowed to change taps until the voltage drops below 50 V.
1. At the top-level menu press the MORE soft key.
2. Press the TAPSET soft key.
3. Use the MODIFY knob to specify a tap (1 through 6) setting.
4. Press the ENTER soft key; you are returned to the output display screen automatically.
LOCKING THE CONTROL PANEL
See "Lock" on page 5-16 for a description of this function.
⇒
MORE
CONFIG
To set the lock code for the first time, follow these steps:
1. At the top-level menu, press the MORE soft key.
2. Press the CONFIG soft key.
3. Press the LOCK soft key.
4. Press the SET CODE soft key.
5. Use the soft keys to enter a five-digit code.
Note: 00000 (five zeros) is not a valid code and will not be accepted.
6. Re-enter the code as instructed.
Either the message "Code Not Accepted—Mismatch" or "Code Accepted' will flash on
the display, and you will be returned to the output display screen.
6-12
⇒
⇒
LOCK
SET CODE
Panel Connection, Controls and Indicators
5700262-P