Chapter 1 Performance Tests

Test 4: Channel 2 Power Measurement

Test 4: Channel 2 Power Measurement
This set of tests verifies the power-measurement specifications of
Channel 2 of the Agilent 53150A/151A/152A Counters.
Equipment Required
Agilent 83650B Synthesizer
Agilent 83051A Microwave Amplifier
Agilent 437B Power Meter
Agilent 8487A/D Power Sensor
Agilent 11730A Power Sensor Cable
Agilent 11667C Power Splitter
Agilent 11904D 2.92 mm (m) to 2.4 mm (f) Adapter
Agilent 11900A 2.4 mm (m) to 2.4 mm (m) Adapter
Agilent 11900B 2.4 mm (f) to 2.4 mm (f) Adapter
2.4 mm (m) to 2.4 mm (m) Cable (2)
Test 4a: 50 MHz to 20 GHz Power Measurement
(–20 dBm)
This test is appropriate for the Agilent 53150A, 53151A, and 53152A
Counters.
Verify that the cable (with Adapter) is still connected to the RF output of
the Agilent 83650B Synthesizer.
Connect the other end of the 2.4 mm cable to the input connector on the
Agilent 83051A Microwave Amplifier.
Connect a 2.4 mm cable from the output of the Microwave Amplifier to the
input connector on the Agilent 11667C Power Splitter.
Verify that the Power Sensor is still connected to the Power Splitter and
that the Power Sensor Cable is still connected between the Power Sensor
and the sensor connector on the Power Meter.
Assembly Level Service Guide
1-25