Note
(Re-)calibration of the keypad
When the lid is closed, all keys are (re-)calibrated (< 5 seconds). During (re-)calibration the
LED is on and the keys cannot be operated.
If one of the keys is pressed for more than 10 seconds, the (re-)calibration of this key begins
which has a duration of less than 10 seconds. Release the key for further operation.
Note
HMI timeout
If no key is pressed for 10 minutes, the display switches to show operation view. If Backlight
is set to Automatic, display backlight goes off automatically 30 seconds after the last
keypress.
Note
Operation does not require opening of the device. This means that the high degree of
protection of IP67 and safety in hazardous areas are guaranteed at all times.
6.5

Initial startup

The first time the device is powered up, you will be prompted to set the language. The device
always starts up showing "Language" in English. When the language has been set, you will
be prompted to set the date and time.
Before using the flow meter for the first time, essential parameters should be considered.
After confirming/changing date and time you can choose to accept the default values or start
the Quick commissioning wizard.
You will be asked if you want to start the "Quick commissioning" wizard. If you choose "Yes"
(recommended), the "Quick commissioning" wizard will start. If you choose "No", you accept
the default values of the device, and the next HMI view will be the operation view 1.
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Commissioning
6.5 Initial startup
53