Modbus communication
A.9 Modbus holding registers tables
Channel 3 discrete input settings
Table A- 58
Channel 3 discrete input settings
Modbus
Data type /
register
Size in bytes
9131
Unsigned / 2
9132
Unsigned / 2
236
Parameter
Description
Input leading
Select the functionality
edge function
for the leading edge of
the active discrete
input signal.
Input trailing
Select the functionality
edge function
for the trailing edge of
the active discrete
input signal.
Default value
Value range /
[units]
Setting options
(units regis-
ter)
0
0: Off
•
6: Reset totalizer 1
•
7: Reset totalizer 2
•
8: Reset totalizer 3
•
9: Reset all totalizers
•
10: Start zero point
•
adjustment
11: Force outputs
•
12: Freeze process
•
values
13: Stop forcing out-
•
puts
14: Unfreeze process
•
values
0
0: Off
•
6: Reset totalizer 1
•
7: Reset totalizer 2
•
8: Reset totalizer 3
•
9: Reset all totalizers
•
10: Start zero point
•
adjustment
11: Force outputs
•
12: Freeze process
•
values
13: Stop forcing out-
•
puts
14: Unfreeze process
•
values
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Access
level
Read /
write
Read /
write