Siemens SITRANS F series Operating Instructions Manual page 103

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F series:
Table of Contents

Advertisement

ID/
Diagnostic
Icons
170
Loop current
172
Transmitter
173
Sensor
174
Display
175
IO
176
Sensor
177
Device is starting
178
Transmitter
179
Alarm class
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Action
Loop current simulated. Disable 'Simulation'
before returning to normal operation
FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
Sensor type incompatible. Please replace
sensor.
Please wait until startup is finished. Startup
time see manual.
FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version.
Alarm class simulated. Disable Simulation
before returning to normal operation.
Diagnosing and troubleshooting
9.4 Fault codes and corrective actions
Comment
Only HART devices
No detailed information avail-
able in view Active diagnostic
events. Icon depends on
simulated alarm class resp.
NAMUR status signal and the
priority in case of multiple
alarm classes simulated.