Modbus communication
A.9 Modbus holding registers tables
Modbus
Data type /
register
Size in bytes
1411
Float / 4
1441
Unsigned / 2
Sensor spacing
Table A- 20
Sensor spacing settings
Modbus
Data type /
register
Size in bytes
1426
Unsigned / 2
1420
Unsigned / 2
164
Parameter
Description
Cable length
Length of the used
sensor cables (One
way). Used for calcu-
lating and compensat-
ing the time delay in
the cables.
Perform fre-
Command to perform a
quency search
frequency sweep on a
path.
Parameter
Description
Recommended
Recommended spac-
space method
ing method between
clamp-on sensors for
path X. Calculated on
request.
Clamp-on rec-
The path the current
ommended
recommended spacing
spacing path X
values were calculated
for. 0 means not calcu-
lated since power up.
Default value
Value range /
[units]
Setting options
(units regis-
ter)
0.0
0.0 - 20.0
[m]
(8556)
0
0: Frequency sweep
done / not running
1: Perform frequency
sweep on path 1
2: Perform frequency
sweep on path 2
3: Perform frequency
sweep on path 3
4: Perform frequency
sweep on path 4
5: Perform frequency
sweep on path 5
Default value
Value range /
[units]
Setting options
(units regis-
ter)
-
0: None (default)
1: Track 1012TN
2: Track 1012TNH
3: Track 992
4: Spacer bar
1012BN
5: Manual
6: Spacing Error
0
0 - 255
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Access
level
Read /
write
Read /
write
Access
level
Read
only
Read
only