Siemens SITRANS F series Operating Instructions Manual page 177

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F series:
Table of Contents

Advertisement

Modbus
Data type /
register
Size in bytes
1065
Unsigned / 2
1070
Float / 4
1072
Float / 4
1074
Float / 4
1076
Float / 4
1078
Float / 4
1080
Float / 4
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Parameter
Description
Multipoint cali-
Status of last per-
bration status
formed Multipoint cali-
bration command.
Calibration
Calibration point 1 for
point 1
undirectional flow or
calibration point 1 for
bidirectional positive
flow.
If 0.0 the calibration
point is not used.
Calibration
Calibration correction
value 1
factor that is associat-
ed to Calibration point
1.
Calibration
Calibration point 2 for
point 2
undirectional flow or
calibration point 2 for
bidirectional positive
flow.
If 0.0 the calibration
point is not used.
Calibration
Calibration correction
value 2
factor that is associat-
ed to Calibration point
2.
Calibration
Calibration point 3 for
point 3
undirectional flow or
calibration point 3 for
bidirectional positive
flow.
If 0.0 the calibration
point is not used.
Calibration
Calibration correction
value 3
factor that is associat-
ed to Calibration point
3.
A.9 Modbus holding registers tables
Default value
Value range /
[units]
Setting options
(units regis-
ter)
0
0: Command suc-
cessfully performed.
1: Sanity check
failed. Not all calibra-
tion points are in
nominal ascending
order.
2: Command not
performed. Calibra-
tion table is protect-
ed.
0.0
0.0 - 100000.0
[m
/s]
3
(7500)
1.0
0.5 - 2.0
0.0
0.0 - 100000.0
[m
/s]
3
(7500)
1.0
0.5 - 2.0
0.0
0.0 - 100000.0
[m
/s]
3
1.0
0.5 - 2.0
Modbus communication
Access
level
Read
only
Read /
write
(Expert)
Read /
write
(Expert)
Read /
write
(Expert)
Read /
write
(Expert)
Read /
write
(Expert)
Read /
write
(Expert)
175