Siemens SITRANS F series Operating Instructions Manual page 159

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F series:
Table of Contents

Advertisement

Modbus
Data type /
register
Size in bytes
1227
Unsigned / 2
1210
Float / 4
1214
Float / 4
1216
Float / 4
1715
Float / 4
1717
Float / 4
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Parameter
Description
Liner material
Liner material.
Liner sound
Liner material sound
velocity
velocity.
Inner pipe
Roughness of inner
roughness
pipe/liner material.
Inner pipe di-
Calculated inner diam-
ameter
eter of pipe.
Calibration
Setup parameter for
reference tem-
specifying the process
perature
medium temperature at
which the flow calibra-
tion of the meter is
performed. The value
is used for compensa-
tion of pipe expansion
due to temperature.
Pipe thermal
Setup parameter for
expansion
specifying the thermal
expansion coefficient
of the pipe material.
The value is used for
compensation of pipe
expansion due to tem-
perature.
A.9 Modbus holding registers tables
Default value
Value range /
[units]
Setting options
(units regis-
ter)
0: Not applicable
1: Custom
2. Cement
3: Coal tar
4: Enamel
5: Glass
6 Plastic
7: Hi_dens_Poly
8: PTFE
9: Rubber
2000.0
200.0 - 4000.0
[m/s]
(7648)
0.0001
0.0 - 0.01
[m]
(8556)
-
-
[m]
(8556)
20.0
-50.0 - +250.0
[°C ]
(7700)
0
-
[1/K]
Modbus communication
Access
level
Read /
write
Read /
write
Read /
write
Read
only
Read /
write
Read /
write
157