Modbus communication
A.9 Modbus holding registers tables
A.9.13

Oil table settings

Table A- 140 Oil table settings (only hydrocarbon applications)
Modbus
Data type /
register
Size in bytes
750
Unsigned / 2
751
Unsigned / 2
752
Unsigned / 2
336
Parameter
Description
Command
Command to update
LiquIdent lookup table
Command
Status of last per-
status
formed Command
Index
Index affected by
Command
Write (ID 1)
Read (ID 2)
Default value
Value range /
[unit]
Setting options
(unit register)
0
0: Command done
/not executing
1: Transfer infor-
mation to oil table
2: Read information
from oil table
3: Perform sanity
check
4: Store table to flash
5: Erase flash
6: Reload latest valid
table from flash
0
0: STATUS_OK -
command executed
successfully
1:
SANITY_CHECK_FA
ILED - sanity check
failed during com-
mand execution
2: MEMORY_FULL -
out of flash storage
space. Erase needed
3:
MEMORY_TIMEOUT
- indicates timeout
during flash
write/erase operation
4:
MEMORY_LOAD_FA
ILED - indicates error
with loading data
from flash
1
1 - 30
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Access
level
Read /
write
Read
only
Read /
write