Siemens SITRANS F series Operating Instructions Manual page 104

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F series:
Table of Contents

Advertisement

Diagnosing and troubleshooting
9.4 Fault codes and corrective actions
ID/
Diagnostic
Icons
181
SensorFlash
183
CH2
184
CH2
185
CH2
186
CH3
187
CH3
188
CH3
189
CH4
190
CH4
102
Action
SensorFlash Chkdsk failed. Start Chkdsk
again. If error still exists replace SensorFlash.
Input current too low. Check wiring and signal
of connected sensor or output source to input
channel.
Input current too high. Check wiring and signal
of connected sensor or output source to input
channel.
External failure. Connected sensor or output to
input channel is out of operation range. Please
check connected sensor or output source.
Input current too low. Check wiring and signal
of connected sensor or output source to input
channel
Input current too high. Check wiring and signal
of connected sensor or output source to input
channel
External failure. Connected sensor or output to
input channel is out of operation range. Please
check connected sensor or output source.
Input current too low. Check wiring and signal
of connected sensor or output source to input
channel
Input current too high. Check wiring and signal
of connected sensor or output source to input
channel
Comment
If Operation mode is config-
ured to Current input.
If Operation mode is config-
ured to Current input
If Operation mode is config-
ured to Current input
If Operation mode is config-
ured to Current input
If Operation mode is config-
ured to Current input
If Operation mode is config-
ured to Current input
If Operation mode is config-
ured to Current input
If Operation mode is config-
ured to Current input
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC