Initial Startup - Siemens SITRANS FS230 Compact Operating Instructions

Ultrasonic flowmeters
Table of Contents

Advertisement

function/parameters. Successful operation of the button is confirmed by a small green LED next
to the display.
Figure 5-1
Note
Calibration of the keypad
When the lid is closed, all buttons are calibrated. During calibration the LED is on and the
buttons cannot be operated.
If one of the buttons is pressed for more than 10 seconds, the calibration of this button begins
which has a duration of less than 10 seconds. Release the button for further operation.
Note
Local display timeout
If no button is pressed for 10 minutes, the display switches to show operation view. If Backlight
is set to Automatic, display backlight goes off automatically 30 seconds after the last button
press.
Note
Operation does not require opening of the device. This means that the high degree of protection
of IP67 and safety in hazardous areas are guaranteed at all times.
5.3

Initial startup

The first time the device is powered up, you will be prompted to set the language. The device
always starts up showing "Language" in English. When the language has been set, you will be
prompted to set the date and time.
You will be asked if you want to start the "Quick commissioning" wizard. If you choose "Yes"
(recommended), the "Quick commissioning" wizard will start. If you choose "No", you accept
the default values of the device, and the next HMI view will be the operation view 1.
SITRANS FS230
Compact Operating Instructions, 12/2019, A5E38755607-AB
Full graphical display
LED (for indication of button operation)
Touch keypad
Local display
Commissioning
5.3 Initial startup
59