Modbus communication
A.9 Modbus holding registers tables
LiquIdent settings
Table A- 34
Liquident settings (only hydrocarbon applications)
Modbus
Data type /
register
Size in bytes
1727
Float / 4
1729
Float / 4
7040
Float / 4
7042
Float / 4
7044
Float / 4
7046
Float / 4
7048
Float / 4
184
Parameter
Description
LiquIdent tem-
Specifies a tempera-
perature slope
ture slope factor used
factor
in connection with the
standard condition
temperature for calcu-
lating liquident.
LiquIdent pres-
Specifies a pressure
sure slope
slope factor used in
factor
connection with the
standard condition
temperature for calcu-
lating liquident.
Upper alarm
Exceeding this limit
limit
causes an alarm.
Upper warning
Exceeding this limit
limit
causes a warning.
Lower warning
Falling below this limit
limit
causes a warning.
Lower alarm
Falling below this limit
limit
causes an alarm.
Hysteresis
Hysteresis that affects
the alarm and warning
limits.
Default value
Value range /
[units]
Setting options
(units regis-
ter)
-4.10
-8.0 - 0.0
[m/K*s]
0.00000435
0.0 - 0.00003
Max. floating
-
point value
[m/s]
(8014)
Max. floating
-
point value
[m/s]
(8014)
Min. floating
-
point value
[m/s]
(8014)
Min, floating
-
point value
[m/s]
(8014)
0.0
-
[m/s]
(8014)
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write