Calibration Factor

Calibration Factor Edit

Thermo Fisher Scientific
may adjust the measured inlet flow rate that is equal at that pump speed.
After pressing the
ENTER
Note Should the temperature and pressure be re-calibrated, the accuracy of
the flow calibration should be verified.
From the Flow Rate Cal screen, press the
forward, until last step (flow calibration point 6, defined flow rate) the
display returns to the Flow Rate Cal screen and the new calibration is
saved only if at least one of the values was edited.
FLOW RATE CAL
ADJ* 188LPM:1.00
The Calibration Factor screen shows the site number selected and allows
the user to edit the calibration factor for that site number. The calibration
factor is the value used to scale the concentration output to agree with a
primary calibration or user defined standard.
In the following screen, the ADR-1500 shows the instrument is operating
for Site 1, and the Calibration Factor is 1.000 which will provide a 1:1
response with the factory calibration. For more information about
calibration, see Chapter 4,
From the Calibrate Menu, press the
key to the Cal Factor screen.
CAL FACTOR
1.000
The Calibration Factor Edit screen allows the user to edit the calibration
factor values. Use the ▲ and ▼ keys to increase or decrease the value by
0.001. The asterisk (*) appears to the left of the field being edited.
From the Cal Factor screen, press the
Press the
key to save the calibration factor associated with the
ENTER
site number and return to Cal Factor screen.
CAL FACTOR
*
1.000
key, the next calibration point is pursued.
ENTER
1
"Calibration and Particle Size
key and scroll the ▲ or ▼
ENTER
1
ENTER
2
MIE ADR-1500 Instruction Manual
Operation
key to navigate
Selection".
key and edit the values.
3-27