This test measures the harmonic d istortion generated in the H P 3562A when a full scale
2-3 1
HARMONIC D ISTORTION
input is present.
The relative ampl itude of all harmonics w i l l be at least 80 dB below the fu ndamental
Specification
ampl itude.
low Distortion Osci llator
Requ i red Test Equi pment
6000 feedthrough termination
S I GNAL ANALYZER
HP 339A
:
0 0
� o o
J
\
6���
T O
TAB
H
R
Figure 2-2 1 Harmonic Distortion Test Setup # 1
A. Connect the test instru ments as shown i n figure 2-21 . Refer to " I n itial Equ ipment
Proced u re
Setu p," paragraph 2-20, for unspecified parameters.