P E RFO RMANCE T ESTS
This test measures the ampl itude accu racy and flatness of the H P 3562A using the ampl itude
2-2 3

AMPLITUDE ACCURACY and FLATNESS

reference of the ac cal ibrator.
If the measurement of a signal is between the BNC center conductor and BNC shell and
Specification
the ampl itude is equal to the range setting, the marker ampl itude reading wi l l not deviate
from the actual signal ampl itude by more than:
+ 27 d BV to - 40 d BV
- 41 dBV to - 51 d BV
If the measurement of a signal incl udes a signal between the BNC shel l and the chassis,
the marker ampl itude read ing will not deviate from the actual signal ampl itude by more
than:
+ 27 dBV to - 40 d BV
- 41 d BV to - 51 d BV
Freq uency Synthesizer
Requ i red Test Equipment
AC Calibrator
BNC Tee
2-34
Range Setti ng
Range Setting
HP 3325A
Fluke 5200A
H P 1 250-0781
Accu racy
± 0.1 5 dB ± 0.01 5 % Range Setting
± 0.25 dB ± 0.025 % Range Setting
Accu racy
± 0.50 dB ± .01 5 % Range Setting
± 0.60 dB ± .025% Range Setting
MOD E L 3562