HP 3562A Service Manual page 31

Dynamic signal analyzer
Hide thumbs Also See for 3562A:
Table of Contents

Advertisement

D. For each of the frequencies listed in table 2-1 perform steps 1 through 7:
1 . Press the HP 3562A keys as fol lows:
RA N G E
F REQ
2. Set the ac calibrator to the signal frequency.
3. Set the freq uency synthesizer to the signal frequency.
4. Set the ac cal ibrator's ampl itude.
5. Press the HP 3562A keys as fol lows:
START
SPCL
MARKER
6. Record the Ya marker reading on the Operational Verification Test Record for CHAN 1 .
7. Record the Yb marker reading on the Operational Verification Test Record for CHAN 2.
Adjustments
If Test Fai ls Check:
Section I I I
Trou bleshooting
Section VI I
To range setting in table
C E N T E R FREQ . . .
MRKR .... .
PEAK
2nd Pass Gain Adjustment
ADC Offset and Reference Adjustment
I nput Flatness Adjustment
I nput Attenuator Adjustments
Cal ibrator Adj ustment
A33, A35 I n put Boards
A32, A34 Analog D igital Converter Boards
A30 Analog Sou rce Board
O P E RAT IONAL V E R I F I CAT ION
To signal freq uency
in table
2-9