Distortion Measurement - HP 333A Operating And Service Manual

Distortion analyzer
Hide thumbs Also See for 333A:
Table of Contents

Advertisement

Model 333A/334A
Section I
SECTION
|
GENERAL
INFORMATION
1-1. DESCRIPTION.
1-2.
The Hewlett-Packard
Models
333A and 334A
Distortion Analyzers are solid state instruments for
measuring distortion on ac voltages.
The
Models
333A and 334A include two control loops that auto-
matically tune both legs of a bridge circuit which re-
jects the fundamental when the rejection circuit is
initially set within the range of the loops.
The 334A
has a high impedance detector which operates
from
550 kHz to greater than 65 MHz and provides the ca-
pability of monitoring the distortion of the amplitude
modulation on an rf carrier.
1-3.
Distortion levels of 0. 1% to 100% full scale are
measured in seven ranges for any fundamental frequen-
cy of 5 Hz to 600 KHz.
Harmonics are indicated upto
3 MHz.
The high sensitivity of these instruments re-
quires only 0. 3V rms for the 100% set level reference.
The distortion characteristics can be monitored atthe
OUTPUT
connectors with an oscilloscope,
a true rms
voltmeter,
ог a wave analyzer.
The instruments are
capable of an isolation voltage of 400 volts above
chassis ground.
1-4.
The voltmeter can be used separately for general
purpose voltage and gain measurements.
It has a
frequency range of b Hz to 3 MHz (20 Hz to 500 kHz
for 300 uV range) and а voltage range of 300uV to
300 V rms full scale,
1-5.
The AM detector included in the Model 334A is
a broadband дс restoring peak detector consisting of
@ semiconductor diode and filter circuit.
AM distor-
. tion levels as low as 0. 3% can be measured on a 3 V
to8 V rms carrier modulated 30% in the standard
broadcast
band.
Distortion less than 1% can be
measured at the same level of the carrier up to 65 Mc.
Table 1-1.
MODEL 333A
DISTORTION
MEASUREMENT
RANGE
Any fundamental frequency,
5 Hz to 600 kHz.
Distortion levels of 0. 1964-1009 are measured
full scale in 7 ranges.
DISTORTION MEASUREMENT
ACCURACY
Harmonic measurement accuracy (full scale)
Fundamental Input Less Than 30 V
100%-0. 3%
0.1%
1-6.
ACCESSORY
FEATURES.
1-7.
The accessory available with the 333A and334A
Distortion Analyzers is a voltage divider probe, -hp-
Model Ко. 10001A.
The features of the probe are:
a.
10 megohms shunted by 10 pF,
giving 10:1
attenuation.
b,
DC to 30 MHz bandwidth.
2% division accuracy.
d.
600 V peak input.
e.
5 ns rise-time.
1-8.
OPTION.
1-9,
Option 01іѕа standard -hp- Model 333A or 334A
with a special meter
and meter
amplifier,
compen-
sated to permit
response
to VU
(volume
units)
characteristics.
1-10 INSTRUMENT IDENTICATION.
1-11.
Hewlett-Packard uses a two-section serial
number.
The first section (prefix) identifies a series
of instruments.
The last section (suffix) identifies a
particular instrument within the series.
На letter is
included with the serial number,
it identifies the coun-
try in which the instrument was manufactured.
Н the
Serial prefix of your instrument differs from the one
on the title page of this manual, a change sheet will
be supplied to make this manual compatable with new -
er instruments or the backdating information in Арреп-
dix C will adapt this manual to earlier instruments.
All correspondence with Hewlett-Packard
should in-
clude the complete serial number.
Specifications
Fundamental Input Greater Than 30 V
100%-0. 3%
10 На
300 ЕН?
30 Hz
300 ЕН?
10 Hz
3 MHz
10 Hz
1.2 MHz
0.1%
Elimination Characteristics:
Fundamental Rejection
> 80 dB
Second Harmonic Accuracy for a
fundamental of:
5 Hz
to
20 Hz:
better than +1 dB
20 Hz
to
20 kHz: better than «0. 6 dB
20 kHz to 100 kHz: better than -1 dB
100 kHz to 300 kHz: better than -2 dB
300 kHz to 600 kHz: better than -3 dB
1-1