Frequency Accuracy Test Setup - HP 3562A Service Manual

Dynamic signal analyzer
Hide thumbs Also See for 3562A:
Table of Contents

Advertisement

FREQUENCY SYNTHESIZER
HP 3325A
00000
88
00000
8 8 8.
00000
0
C!J
00000
Figure 2-1 6 Frequency Accuracy Test Setup
A. Connect the test equipment as shown in figure 2-1 6. Refer to "I nitial Equipment Setu p,"
Procedu re
paragraph 2-20, for unspecified parameters.
B. Set the test instruments initial ly as fol lows:
Frequency Synthesizer
Frequency
Ampl itude
Fu nction
C. Press the H P 3562A keys as fo l lows:
PRESET
CAL
RA N G E
F REQ
AVG
START
x
HP 3562A
8 · .
r-
8NC CABLE
8 1 2 0 - 1 8 4 0
99 kHz
1 Vrms
Sine Wave
R E S E T
S I NG L E
CAL
d BVrms
C E N T E R FREQ
o
FREQ SPAN
STA BLE
2
P E R FORMANCE T ESTS
@
HP
1 1 0 48C
50 OHM FEED THROUGH
T E R M I N A T I ON
kHz
kHz
9 9
E NT E R
. 5
2-51