HP 3562A Service Manual page 48

Dynamic signal analyzer
Hide thumbs Also See for 3562A:
Table of Contents

Advertisement

O P E RAT IONA L V E R I F I CATION
MEAS
MODE
SOU RCE
U N ITS
FREQ
C. When the sweep is com plete perform steps 1 and 2 :
START
1 . Press the H P 3562A keys a s fol lows:
SCALE
2 . If the trace i s between the 9 dB and the 1 1 dB lim its, check PASS o n the
Operation Verification Test Record for the 0 to 65 kHz span.
D. Press the H P 3562A keys as follows:
FREQ
E. When the sweep is complete perform stage 1 and 2:
START
1 . Press the H P 3562A keys as follows:
SCALE
2. If the trace is between the 8.5 dB and the 1 1 dB lim its, check PASS on the Opera­
tion Verification Test Record for the 65 kHz to 1 00 kHz span.
Adj ustments
If Test Fai ls Check:
Troubleshooting
Section V I I I
2-26
SWE PT
S I N E
SOURC E L EVE L . . . .
P SPEC
U N ITS
STOP
FREQ
Y FIXD
SCALE
START
FREQ
Y FIXD
SCALE
None
A30 A nalog Sou rce Board
MOD E L 3562
L I N EA R
SWE E P
V
VO LTS
4.47
RMS
VOLTS
kHz
65
9,1 1 d B
kHz
65
dB
8.5, 1 1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents