PE RFO RMAN C E TE STS
Range
Setting
8 d BV
8 d BV
- 1 1 d BV
- 1 3 d BV
- 1 3 d BV
- 1 3 dBV
- 1 3 d BV
- 27 d BV
- 27 d BV
Ampl itude
1 0.00 Vrms
1 .000 Vrms
1 00.0 mVrms
1 0.00 mVrms
3.1 623 mVrms
1 .000 mVrms
1 0.00 Vrms
1 .000 Vrms
1 00.0 mVrms
1 0.00 mVrms
3.1 623 mVrms
2 -96
1 .000 mVrms
2-23
Amplitude Accuracy and Flatness
Measurement Three
BNC center conductor grou nded
Specification
S ignal
Frequ ency
Lower Limit
1 k H z
7.499 d BV
9 9 k H z
7.499 d BV
1 k H z
- 1 1 .50 d BV
1 k H z
- 1 3.50 d BV
5 0 k H z
- 1 3.50 d BV
9 0 k H z
- 1 3 .50 d BV
9 9 k H z
- 1 3.50 d BV
1 k H z
- 27.50 d BV
9 9 k H z
- 27 .50 d BV
2-24

Amplitude Linearity

Signal Frequency
1 0 k H z
=
BNC shell grounded
S pecification
Upper Limit
1 0.1 8 Vrms
1 .01 9 Vrms
1 03.2 mVrms
1 1 .67 mVrms
4.71 7 mVrms
2.51 7 mVrms
BNC center conductor grou nded
1 0.59 Vrms
1 .061 Vrms
1 07.4 mVrms
1 2.09 mVrms
4.850 mVrms
2.559 mVrms
Upper Limit
C H AN 1
8.501 d BV
8.501 d BV
- 1 0.50 d BV
- 1 2.50 d BV
- 1 2.50 d BV
- 1 2 .50 d BV
- 1 2.50 d BV
- 26.50 d BV
- 26.50 d BV
Range Setting
1 0 Vrms
=
Measured Value
Lower Limit
C H AN 1
9.827 Vrms
981 .4 mVrms
96.79 mVrms
8.329 mVrms
1 .608 mVrms
- 51 7.1 ftVrms
9.439 Vrms
942.6 mVrms
92.91 mVrms
7.941 mVrms
1 .485 mVrms
- 555.9 ftVrms
MOD E L
3562
Measured Value
C H AN 2
CHA N 2

Hide quick links: