Amplitude Accuracy And Flatness Test Setup - HP 3562A Service Manual

Dynamic signal analyzer
Hide thumbs Also See for 3562A:
Table of Contents

Advertisement

i M OD E L 3562A
AC CALI BRATOR
FLUKE 5 2 0 0 A
REAR PANEL V I EW
FRONT PANEL V I E W
0 0 0 0
Figure 2-1 2 Amplitude Accuracy and Flatness Test Setup
A. Connect the test instru ments as shown in figure 2-1 2. Refer to "I nitial Equipment Setup,"
Proced u re
paragraph 2-20, for u nspecified parameters.
B. Set the test instruments initial ly as fol lows:
Freq uency Synthesizer
Ampl itude
Frequency
Function
AC Cali brator
Phase Lock
Sense
Mode
Frequency
Ampl itude
PHASE
LOCK
INPUT
D
(i)
(i)
0
0
11
o •
OUTPUT
0.5 Vrms
1 kHz
S ine Wave
O N
I NT E RNAL
O P E R
1 kHz
2.81 84 Vrms
FREQUENCY SYNTHESIZER
HP 3325A
o 8
88
00000
0 0 0 .
00000
00000
00000
BNC CABLE
8 1 2 0 - 1 8 4 0
;rnOoo[gJ
HP 3562A
@
8NC/OUAL 8ANANA
1 1 0 0 1 - 60 0 0 1
BANANA/BANANA
P E R FORMANCE T ESTS
0 ·
ill
8NC C A B L E
8 1 20 - 1 83 8
'--- --
TEE
1250-07 8 1
'-
F TO
F BARREL
1 2 5 0
- 0 0 8 0
2-35