. OD E L 3562A
2- 1 6
OPERATIONAL VERIFICATION TEST RECORD
Range
Setti n'g
- 51 dBV
Range
Setting
Frequency
9 dBV
9 dBV
o
dBV
o
dBV
- 1 3 dBV
- 1 3 dBV
Part
Range
Setting
- 49 dBV
o
dBV
1 0 dBV
S ignal Frequency
99,000 H z

2-7 Self Test

2-8 DC Offset

Measured Value
C H AN 1
2-9 Amp litude Accuracy and flatness
CHAN 1 and CHAN 2 Floating
Spec ification
Signal
Lower L i m it
1 kHz
8.849 dBV
99 kHz
8.849 dBV
1 kHz
- 0.1 51 3 dBV
99 kHz
- 0.1 51 3 dBV
1 kHz
- 1 3 .1 5 dBV
99 kHz
- 1 3.1 5 dBV

2-10 Amplitude and Phase Match

PASS
Ampl itude
Specification
± 0.1 dB
1
2
± 0.1 dB
3
± 0.1 dB

2-11 frequency Accuracy

Specification
Lower Limit U pper L i m it
98.996 kHz
O P E RATIONAL V E R I FICAT ION
PASS
Specification
CHAN 2
< - 71 dBV
Measured Value
Upper Limit
CHAN 1
9.1 51 dBV
9.1 51 dBV
0.1 51 3 dBV
0.1 51 3 dBV
- 1 2.85 dBV
- 1 2.85 dBV
PASS
Part
4
5
6
Measured Value
99.004 kHz
CHAN 2
Phase
Specification
± 0.5 °
± 0.5 °
± 0.8 °