HP 3562A Service Manual page 553

Dynamic signal analyzer
Hide thumbs Also See for 3562A:
Table of Contents

Advertisement

MO D E L 3S62A
8-11 2
H P 1 345A DISPLAY
The H P 1 345A d is p l ay is a stand alone d igital d is p l ay. Use the H P 1 345A service m a n u a l
i n c l uded w i t h the i n stru ment to service the d is p l ay. T o verify the d is p l ay is fai l i ng, set
j u mper A1 7 W1 to the test (T) position. The pattern displayed shou ld be the same as shown
in fi g u re D-1 . To re move the d is p l ay from the i n stru ment perform steps 1 through 8:
1 .
Di scon nect the m a i n power cord from the rear panel and rem ove the top cove r.
2 .
Remove the i nternal s h ie l d covering the d i s p l ay u n it.
3.
Remove the trim strips from the front frame.
4.
Remove the front frame and s i d e panel screws as shown in figu re D-2 .
5 .
Rem ove the sc rews that attach the d i splay to the i nstrum ent as shown i n f i g u re D-2 .
6.
Remove the two sc rews attac h i ng the d i splay adjustme nts to the rear panel of the
i n strum ent.
7.
Di sconnect the d is p l ay power cable (W1 1 ).
8.
Discon nect the d isplay i nterface cables (W1 4, W1 90, W1 91 , W1 92). The d is p l ay can
now be pu l led from the front of the i n stru ment.
0-1
Figure
Oisplcll y of Verification Pattern
S E RV I C E
8-1 03

Hide quick links: