P E R FORMANCE T E STS
C. Press the H P 3562A keys as fol lows:
PRESET
CA l
I N P U T
C O U P LE
W I N DOW
AVG
U N ITS
A & B
Table 2-6 Amplitude Accuracy and Flatness
Measurement One
H P 3562A
Range
Setting
9 d BVrms
9 d BVrms
- 1 3 d BVrms
- 1 3 d BVrms
- 1 3 d BVrms
- 1 3 d BVrms
- 23 d BVrms
- 23 d BVrms
- 26 d BVrms
- 21 d BVrms
- 1 7 d BVrms
- 1 4 d BVrms
- 1 1 d BVrms
2-36
R E S E T
S I NG LE
CAL
G ROUND
CHAN 1
G ROUND
CHAN 2
F LAT TO P
STA BLE
4
P SPEC
U N ITS
BNC shel l grou nded
Signal
AC Cali brator
Frequ ency
Amplitude
1 kHz
2.81 84 Vrms
99 kHz
2.81 84 Vrms
1 kHz
.22387 Vrms
50 kHz
.22387 Vrms
90 kHz
.22387 Vrms
99 kHz
.22387 Vrms
1 kHz
70.795 mVrms
99 kHz
70.795 mVrms
1 kHz
50.1 1 9 mVrms
1 kHz
89.1 25 mVrms
1 k H z
.1 41 25 Vrms
1 kHz
.1 9953 Vrms
1 kHz
.281 84 Vrms
MOD E L 3562
E N T E R
VO LTS
RMS
VOLTS
Specifi cation
Lower Limit
Upper Limit
8.849 dBV
9.1 51 d BV
8.849 d BV
9.1 51 d BV
- 1 3.1 5 d BV
- 1 2.85 d BV
- 1 3.1 5 d BV
- 1 2.85 d BV
- 1 3.1 5 d BV
- 1 2.85 d BV
- 1 3.1 5 d BV
- 1 2.85 d BV
- 23.1 5 d BV
- 22.85 d BV
- 22. 8 5 d BV
- 23.1 5 dBV
- 26.1 5 d BV
- 25.85 d BV
- 21 .1 5 d BV
- 20.85 d BV
- 1 7.1 5 d BV
- 1 6.85 d BV
- 1 4.1 5 d BV
- 1 3.85 d BV
- 1 1 .1 5 d BV
- 1 0.85 d BV