V
Section
VOLTMETER CALIBRATOR
hp738BR
0
©
@
0
'"
'"
0
))00_-'-'44'
Figure 5-1. Accuracy. Linearity. and DC Output Test Setup

5-9. FREQUENCY RESPONSE CHECK.

Connect 50n feedthru termination
directly to 3400A INPUT to elimi­
nate loss in output cable at high
frequency.
a.
Connect test setup illustrate d i n Figure 5-2.
b. Set Model 3400A RANGE switch and Test
Oscillator output atienuator to 1 volt position .
c . Set Test Oscillator for 400 H z output and ad­
just output ampli t ude for the reading obtained
in Paragraph 5-8 step d on the Model 3400A.
d . Switch the Test Oscillator monitor switch to
expand mode and set a convenient reference
level.
e. Adjust test oscillator output frequency to
values listed under ' ' frequency''
5-3; adjust oscillator
maintain reference level set in step d .
Model 3400A does not indicate within values
under "meter reading" in Table 5-3. perform
high frequency calibration procedures. Para­
graph 5-25 .
TEST OSCILLATOR
__ '_,41>0
uoo
5-2
hp3400A
RMS VOL TMETER
o O Ii>
0
o.c, OUT JACK
ON REAR
in Table
output voltage t o
If
©
©
hp 652A
@ m
Figure 5-2.

Frequency Response Test Setup

o G
I
I
TA
@
0
TO
Table 5-3. Frequency Response Check
FREQUENCY
15 Hz
4 5 Hz
100 Hz
900 kHz
1 . 2 MHz
1. 8 MHz
2. 2 MHz
2. 8 MHz
3 . 2 MHz
9. 8 MHz
5-10 . INPUT IMPEDANCE CHE C K .
5-ll. RESISTANCE CHECK.
a.
Connect the Test Oscillator. 50 ohm feedthru
and -hp- Model 3400A to position A in Figure
5-3.
b. Set 3400A t o 1 volt range and Test Oscillator
to 400 Hz.
c. Adjust Test Oscillator output for 1 volt 'in­
dication on Model 3400A .
�on
FEED-TttRU
TERMINATION
h PI I 0488
- - - - - - -
Model 3400A
L V L M
0 T ETER
@
hp3440A
hp3443A
HIGH GAIN/
AUTO RANGE
.
v
UNIT
METER READING
0 . 95 to 1 . 0 5
0. 95 to 1 . 05
0 . 99 to 1 . 01
0 . 99 to 1 . 01
0. 98 to 1 . 02
0. 98 to 1 . 02
0. 97 to 1. 03
0. 97 to 1 . 03
0. 95 to 1 . 0 5