Instrument Accuracy Test

Specification

Electrical Characteristics
Accuracy
1. This performance limit does not include the corresponding sensor power
linearity specification.

Description

The power meter accuracy is verified for various power inputs. The range
calibrator is switched to provide these reference inputs.

Equipment

Range Calibrator ................................... HP 11683A
Power Sensor Cable .............................. HP 11730A

Test Setup

Power Meter
Figure 2-2: Instrument Accuracy Test Setup
HP EPM-441A/442A Service Guide
Performance Limits
± 0.5% or ±0.02 dB
CHANNEL A
HP 11730A
Performance Tests
Instrument Accuracy Test
1
Range
Calibrator
POWER
METER
2-7