1. Test Port Output Frequency Range and Accuracy

Specifications
Required Equipment
kequency Counter (30 kHz to 500 MHz) . . . . . . . . . . . . . . . . . . . . HP 5350B/51B/52B
Cable, 500 Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP P/N 81204781
Adapter, APC-3.5 (f) to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 1250-1745
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 11524A
Adapter, Type-N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 1250-1477
Additional equipment needed for an HP 8753E with Option 075
Minimum Loss Pad, 5OQ to 75Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 11852B
Perform this test to verify the frequency accuracy of the HP 8753E over its
entire operating frequency range.
Performance Tests
Frequency Range Frequency Accuracy1
30
1 At 25O C f5* C.
2 Only for analyzers with Option 006 - 30 kHz to
minutes
I