Specifications

Model
Color
Illumination/
viewing system
Detector
Spectral separation
device
Wavelength range
Wavelength pitch
Half bandwidth
Measurement range
Light source
Measurement/
illumination area
Repeatability
Inter-instrument
agreement
Observer
Illuminant
Displayed data
Colorimetric data
Indexes
Color-difference
equation
Standard compliance
Gloss
Measurement
geometry
Light source
Detector
Measurement range
Measurement area
Repeatability
Inter-instrument
agreement
Standard compliance
Measurement time
Minimum measurement interval
Battery performance
121
Spectrophotometer CM-25cG
45°c: 0°
Dual 40-element silicon photodiode arrays
Planar diffraction grating
360-740 nm
10 nm
Approx. 10 nm
0-175%; Output/display resolution: 0.01%
Pulsed xenon lamp
MAV: Ø8 mm/12×16 mm, SAV: Ø3 mm/12×16 mm
Chromaticity value: Standard deviation within ΔE*ab 0.04
(When a white calibration plate is measured 30 times at 10-second
intervals after white calibration)
Within ΔE*ab 0.15 (MAV)
(Average for 12 BCRA Series II color tiles compared to values measured
with a master body under Konica Minolta standard measurement
conditions)
2° or 10° Standard Observer
A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12, ID50, ID65, User illuminant
(simultaneous evaluation with two light sources possible)
Spectral values/graph, colorimetric values/graph,
color-difference values/graph, pass/fail judgement, pseudocolor
L*a*b*, L*C*h, Hunter Lab, Yxy, XYZ, and color differences in these
spaces; Munsell
MI, WI (ASTM E313), YI (ASTM E313, ASTM D1925),
ISO Brightness (ISO2470), WI/Tint (CIE)
ΔE*ab (CIE1976), ΔE*94 (CIE1994), ΔE00 (CIE DE2000), CMC (l:c),
ΔE (Hunter), ΔE99o (DIN99o)
CIE No.15, ISO 7724/1, ASTM E179, DIN 5033 part7, JIS Z8722
60°
LED
Silicon photo diode
0-200 GU; Output/display resolution: 0.01 GU
MAV: Ø10 mm, SAV: Ø3 mm
0-10 GU: 0.1 GU
10-100 GU: 0.2 GU
>100 GU: 0.2% of displayed value
(Standard deviation. Under KONICA MINOLTA standard conditions)
0-10 GU: ±0.2 GU
10-100 GU: ±0.5 GU
(MAV. Compared to values measured with a master body under KONICA
MINOLTA standard conditions)
JIS Z8741, JIS K5600, ISO 2813, ISO 7668, ASTM D523-08,
ASTM D2457-13, DIN 67530
Approx. 1 second (to data display/output)
Approx. 2 seconds
Approx. 3,000 measurements/charge
(Stand-alone measurement at 10-second intervals at 23°C)
Approx. 1,000 measurements/charge
(When using Bluetooth® communication)