Specifications

Illumination/
Reflectance (*1)
viewing system
Detector
Silicon photodiode array with flat holographic grating
Spectral separation
Diffraction grating
device
Wavelength range
400 to 740 nm
Wavelength pitch
10 nm
Approx. 14 nm average
Half bandwidth
Measuring range
0 to 200%; Resolution: 0.001%
Light source
Pulsed xenon arc lamp
Minimum
3 seconds
measurement
interval
Measurement/
Reflectance
illumination area
Spectral reflectance: Standard deviation within 0.05%
Colorimetric values: Standard deviation within ∆E*
Repeatability(*2)
(When a white calibration plate is measured 30 times at 10-second intervals after white
calibration)
Within ∆E*
Inter-instrument
agreement (*2)
measured with a Konica Minolta master body)
Interface
USB 1.1
Power
AC 100 to 240 V 50/60 Hz 25 VA (using included AC adapter)
13 to 33 ° C, relative humidity 80% or less (at 33°C) with no condensation
Operation
temperature/
humidity range
0 to 40 ° C, relative humidity 80% or less (at 35 ° C) with no condensation
Storage
temperature/
humidity range
Size (W x H x D)
271 x 274 x 500 mm (10-11/16 x 10-3/4 x 19-11/16 in.)
Weight
Approx.16.1 kg (35.5 lb.)
*1 This instrument is designed only for measuring reflected color.
*2 When using Target Mask (SAV) under Konica Minolta standard measurement conditions.
The specifications and appearance shown herein are subject to change without notice.
di:8 ° , de:8 ° (diffuse illumination/8 ° viewing angle)
SCI (specular component included)/SCE (specular component excluded)
switchable
Conforms to CIE No.15(2004), ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7
and JIS Z 8722 condition c standard.
USAV : 1x3 mm measurement / 3x5 mm illumination
SAV
: 1x3 mm measurement / 5x7 mm illumination
0.2 (Based on average for 12 BCRA Series II color tiles; Compared to values