Specifications

di: 8°, de: 8° (diffused illumination, 8° viewing angle)
SCI (specular component included) / SCE (specular component excluded) switchable
Conforms to CIE No.15 (2004), ISO7724/1, ASTM E1164, DIN 5033 Teil7, JIS Z 8722
Condition c standard
Illumination/
viewing
di: 0°, de: 0° (diffused illumination, 0°
system
viewing)
Conforms to CIE No.15 (2004), ASTM E1164,
DIN 5033 Teil7, JIS Z 8722 Condition g
standard
Size of integrating
ø152 mm (6 inches)
sphere
Detector
Dual 40-element silicon photodiode arrays
Spectral
Diffraction grating
separation device
Wavelength
360 to 740 nm
range
Wavelength pitch 10 nm
Half bandwidth
Approx. 10 nm
Reflectance range 0 to 200%; Resolution: 0.01%
Light source
Pulsed xenon lamps × 3 (2 with UV cut filters) Pulsed xenon lamp × 1
Illumination area
(mm)
Measurement
area (mm)
Colorimetric values: Standard deviation
within ΔE*ab 0.02
Spectral reflectance: Standard deviation
Repeatability
within 0.1%
(When a white calibration plate is measured
30 times at 10-second intervals after white
calibration)
Within ΔE*ab 0.12 (Based on average for 12
BCRA Series II color tiles; LAV/SCI. Compared
Inter-instrument
to values measured with a master body under
agreement
Konica Minolta standard measurement
conditions)
100% / 0% / Adjusted (Instantaneous
numerical adjustment of UV with no
UV setting
mechanical filter movement required)
400 nm and 420 nm UV cutoff filters
*1 Numerical adjustment of UV requires UV Adjustment Software (included with optional SpectraMagic NX Pro Ver. 3.2 or later).
CM-36dG
LAV
LMAV
MAV
ø30
ø20
ø11
ø25.4
ø16
ø8
SAV
Transmittance
LAV
ø7
ø24
ø30
ø4
ø17
ø25.4
Colorimetric values: Standard deviation
within ΔE*ab 0.03
Spectral reflectance: Standard deviation
within 0.1%
(When a white calibration plate is
measured 30 times at 10-second
intervals after white calibration)
Within ΔE*ab 0.15 (Based on average for
12 BCRA Series II color tiles; LAV/SCI.
Compared to values measured with a
master body under Konica Minolta
standard measurement conditions)
No adjustment function (UV100%)
*1
;
39
CM-36d
MAV
SAV
ø11
ø7
ø8
ø4