Microsoft Corporation in the United States and other countries. Notes on this Manual • Copying or reproduction of all or part of the contents of this manual without KONICA MINOLTA’s permis- sion is strictly prohibited.
Should liquid or metal objects AC adapter other than those specified by enter the instrument, turn the power KONICA MINOLTA is used, it may result in OFF immediately, disconnect the damage to the unit, fire or electric shock. AC adapter plug from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
Keep all packing materials (cardboard box, cushioning material, plastic bags, etc.) in a safe place. The CM-3600A is a precision measuring instrument. They can be used to protect the instrument from impact and vibration during shipment to Konica Minolta for maintenance.
Power Source • Make sure that the power switch is set to OFF (" I ") when the CM-3600A is not in use. • Always use the AC adapter supplied as a standard accessory (AC-A308) and connect it to an AC outlet of the rated voltage and frequency.
Page 6
• The specimen measuring port and integrating sphere are extremely precise components, and great care should be taken to prevent them getting dirty or exposing them to impacts. When the CM-3600A is not in use, be sure to attach a target mask to the measuring port to prevent entry of foreign matter.
And store them in the accessory case (CM-A215). • Take care not to leave the CM-3600A for a long period of time with a target mask attached. The sample holder may stick to the target mask.
Notes on Storage..........5 Notes on Cleaning ..........5 Disposal Method ..........5 CONTENTS ..........6 Using the CM-3600A Cleaning the CM-3600A and Accessories ..24 Zero Calibration Box and White Calibration Plate ..24 Standard Accessories ......8 Target Mask ............24 Optional Accessories ......9 Inside Integrating Sphere .........
Used to perform zero calibration for measurement of reflectance and to perform haze measurement at measurement of transmittance. AC Adapter AC-A308 Used to supply power from an AC outlet to the CM-3600A. 50/60 Hz Input: 100 to 240 V Output: 8 V 1.5 A Plug design:...
SpectraMagic™ NX CM-S100w This software provides various functions (e.g., data processing and file management) and allows the user to operate the CM-3600A using a personal computer. Transmittance Specimen Holder CM-A96 Used to hold the specimen for measurement of transmittance. It can hold specimens up to 22.5 mm (7/8 in.) thick.
⑧ USB connection terminal ..Used to connect the instrument to a PC with the supplied USB cable (IF-A21 or IF-A22). ⑨ AC adapter input socket ..Used to connect the AC adapter supplied with the CM-3600A.
Measurement Procedure ○ This manual explains how to prepare the CM-3600A and how to set a specimen. ○ The CM-3600A is controlled by a PC to perform measurements. ○ For a description of measuring method using SpectraMagic™ NX (optional), refer to the SpectraMagic™ NX instruction manual.
• The USB communication port of the instrument conforms to USB 1.1. ○ To connect the instrument to a PC, you need to install the USB driver dedicated to the CM-3600A. Install the USB driver supplied with the software that enables connection and operation of the instrument.
Failure to do so may damage the CM3600A or the AC adapter, causing a fire or electric shock. If the CM-3600A will not be used for a long time, disconnect the AC adapter from the AC outlet. Accumulated dirt or water on the prongs of the AC adapter's plug may cause a fire and should be removed.
Turning Power ON and OFF WARNING The CM-3600A should not be operated if the CM-3600A or the AC adapter is damaged, or smoke or strange odors occur. Doing so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
○ When not using the CM-3600A, attach one of the target masks or the protective cap to prevent dust entering the integrating sphere. ○ Take care not to leave the CM-3600A for a long period of time with a target mask attached. The sample holder may stick to the target mask.
Do not perform measurement with the specimen measuring port directed towards your face. Doing so may cause damage to your eyes . Be careful around openings in the CM-3600A. Failure to do so may result in fingers being trapped causing injury.
Do not perform measurement with the specimen measuring port directed towards your face. Doing so may cause damage to your eyes . Be careful around openings in the CM-3600A. Failure to do so may result in fingers being trapped causing injury.
Attaching the Transmittance Zero Calibration Plate (Optional Accessory) The Transmittance Zero Calibration Plate is used for performing zero calibration for transmittance measurements. ○ Before starting transmittance calibration and measurements, attach the LAV (ø25.4 mm) target mask and ○ position the white calibration plate at the reflectance measuring port. ○...
Performing 100% calibration Two methods for performing 100% calibration for transmittance measurements are possible: • 100% calibration to air: When the specimen to be measured is in sheet or solid form, 100% calibration should be performed with the transmittance chamber empty. •...
Doing so may cause a fire. Do not disassemble or modify the CM-3600A. Doing so may cause a fire or electric shock. The CM-3600A should not be operated if it is damaged, or smoke or strange odors occur. Doing so may result in a fire.
⿎ Reflectance measurement ○ Make sure that there is nothing in the transmittance chamber. ○ Before starting reflectance measurements, attach the target mask to be used for measurements. ○ Before starting reflectance measurements, set the measurement area, specular component, and UV light quantity using the software.
Doing so may cause a fire or electric shock. The CM-3600A should not be operated if it is damaged, or if smoke or strange odors occur. Doing so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
The following error messages may appear when you use SpectraMagic® NX CM-S100w, optional color management software, to control the instrument. If you see an error message, follow the instructions shown in the table below. If a problem persists, contact the nearest Konica Minolta-authorized service facility. Error Messege...
Page 28
COM port and • COM port setting communication settings. is incorrect. • Communication settings are incorrect. Detected instrument The instrument is incorrect. Connect CM-3600A. disagrees with connected — one.\nConnect again. File is different to The measurement Adapt the actual current status.
TROUBLESHOOTING GUIDE If a problem occurs with the Spectrophotometer, please check the following points before requesting service. If the problem continues to occur even after the suggested corrective actions have been taken, contact the nearest Konica Minolta-authorized service facility. Refer Condition...
The flow of measurement is shown below. The geometry of the CM-3600A conforms to CIE No.15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 condition c (diffused illumination/perpendicular viewing system) standards, and offers both di:8° (SCI: specular component included ;...
⿎ Target Mask Since the CM-3600A has no target mask detection function, it is not possible for the instrument to determine which target mask has been attached. Thus, when switching from one target mask to another, be sure that the intended target mask is attached and the instrument is properly set from the software.
SCE need to be switched. In addition, it is not possible to start measurement until switching is completed. The CM-3600A has eliminated mechanical switching and enables acquisition of SCI and SCE data by performing calculations with the measured data obtained using two light sources.
⿎ WHEN FLUORESCENT CALIBRATION IS NOT PERFORMED The UV full xenon lamp of the CM-3600A has a spectral distribution similar to that of the D65 light source. Therefore, when high-accuracy fluorescent measurements are not required, normal measurement of reflectance can be used to measure fluorescent reflectance.
Specifications Model CM-3600A Reflectance; di:8°, de:8° (diffused illumination, 8-degree viewing), equipped with simultaneous measurement of SCI (specular component included) / SCE Illumination/ (specular component excluded) Conforms to CIE No.15, lSO 7724/1, ASTM observation E 1164, DIN 5033 Teil 7 and JIS Z 8722 condition c standard. Transmittance system : di:0°, de:0°...
Page 39
< CAUTION > KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE MISUSE, MISHANDLING, UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT, OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE...