The signal patterns for the buzzer consist of 2 identical pulse sequences, repeated every 10 seconds. Each
pulse sequence consists of 5 individual pulses (see Table of Signal Patterns).
Table of Signal Patterns:
Buzzer pattern
Pulse duration / ms
for one pulse
(= buzzer on)
sequence
A
B
C
Flashing pattern
Flashing
for
frequency / Hz
red LED
A
B
C
D

6.1.3 Deactivating the signal functions

The signal functions can be deactivated by means of the switches
board. To see where the switches are located on the PCB, consult the illustrations in Section "10.1 Arrange-
ment of the PCBs" on page 76.
Coding switch
S2
S3
S4
S5
Tip:
For fault-finding purposes it can be useful to open coding switch
completion of the leak test.
Pulse interval/ ms
(= buzzer off)
150
100
150
150
On time / ms
1.43
200
1.43
400
2.5
80
2.5
240
Open
Power supply for power failure alarm
deactivated.
Buzzer deactivated.
Alarm suppression during warming-up
phase deactivated
(i.e. alarm function activated).
Activates program for checking status
indicator
Pulse interval
between 3rd and
4th pulse / ms
500
10
320
0
300
Off time / ms
500
300
320
160
,
,
S2
S3
Power supply for power failure alarm
activated.
Buzzer activated.
Alarm suppression during warming-up
phase activated
(i.e. alarm function deactivated).
Normal operation
. This enables the alarms immediately on
S4
Graphic representation
— — —
— — —
— —
———
——
Graphic representation
——
——
— —
and
on the printed circuit
S4
S5
Closed