Scpi Status Model - Keithley 3390 User Manual

Arbitrary waveform generator
Hide thumbs Also See for 3390:
Table of Contents

Advertisement

Appendix A: SCPI Command Reference

SCPI status model

The SCPI status model shows the structure of the various registers used by the Model 3390
(Figure
manipulate events. The core of the model is the Status Byte Register.
Figure A-1:
Model 3390 status model
Questionable Data Register
C
Voltage overload
0
1
2
3
4
Loop unlocked
5
6
7
Calibration error
8
External reference
9
10
11
12
13
14
15
STAT:QUES:COND?
STAT:QUES:EVEN?
Standard Event Register
EV
Operation complete
0
1
Query error
2
Device error
3
Execution error
4
Command error
5
6
Power on
7
*ESR? *ESE<value>
A-6
A-1). These registers record conditions of the instrument and allow you to monitor and
EV
EN
0
< 1 >
1
< 2 >
2
< 4 >
3
< 8 >
4
< 16 >
5
< 32 >
Logic
6
< 64 >
7
< 128 >
8
< 256 >
9
< 512 >
10
< 1024 >
11
< 2048 >
12
< 4096 >
13
< 8192 >
14
< 16384 >
15
NOT USED
STAT:QUES:ENAB<value>
STAT:QUES:ENAB?
Output buffer
EN
< 1 >
< 2 >
< 4 >
Logic
< 8 >
OR
< 16 >
< 32 >
< 64 >
< 128 >
*ESE?
Return to
Model 3390 Arbitrary Waveform Generator User's Manual
OR
Error Queue
SYST:ERR?
Section Topics
Status Byte Register
C
EN
0
< 1 >
1
< 2 >
2
< 4 >
3
< 8 >
4
< 16 >
5
< 32 >
6
RQS
7
< 128 >
*STB?
*SRE<value>
*SRE?
C = Condition Register
EV = Event Register
EN = Enable Register
3390-900-01 Rev. C / January 2009
Logic
OR

Advertisement

Table of Contents
loading

Table of Contents