Kv, 20A, High Voltage Medium Current Configuration - Keysight Technologies B1505A Configuration Manual

Power device analyzer/curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Connection and Ordering Examples
Package Device Measurement Configuration Examples

3 kV, 20A, High Voltage Medium Current Configuration

Figure 7-8
Connection example
16493G-001 Digital I/O cable (Included in N1266A)
B1505A
16493T (Included in the HVSMU)
HVSMU
For Drain
16493S
(Included in the HCSMU)
F
For Drain
HCSMU
S
16494A (Included in the MCSMU)
F
MCSMU
For N1266A
S
F
MCSMU
For N1266A
S
F
MCSMU
For Gate
S
16493L (Included in B1505A)
GNDU
16493J-001 Interlock cable (included in B1505A)
Table 7-8
Ordering example
Model/Option
Quantity
B1505A
B1512A-FG
B1513C-FG
B1514A-FG
16493L
16493L-001 or
002
16493T
16493T-001 or
002
N1254A
7-16
Device Type: Discrete MOSFET (3 terminals)
Key Specification
Max voltage/current: 3 kV/20A
High Voltage Medium Current capability: 2200 V/1.1 A, 1500 V/2.5 A
Module Selector for multi-parameter automated measurement function
Input
Input
HVSMU
N1266A
Vcontrol F
Vcontrol S
Icontrol F
Icontrol S
GNDU
1
Power Device Analyzer/Curve Tracer mainframe
1
High Current Source Monitor Unit, 20 A/20 V(Pulsed); 1 A/40 V(DC)
1
High Voltage Source Monitor Unit, 3000 V/4 mA (Pulsed & DC)
3
Medium Current Source Monitor Unit, 1 A/30 V(Pulsed), 100 mA/30 V(DC)
1
Ground Unit Cable (Triaxial)
1
Ground Unit Cable (Triaxial, 1.5m or 3.0m)
1
High Voltage Source Monitor Unit Cable
1
High Voltage Source Monitor Unit Cable (1.5m or 3.0m)
1
Accessories for instruments and fixtures
16493G-001 Digital I/O cable (Included in N1259A-300)
Output
16493T
Output
High
N1254A-104 TRX(J) to BNC(P) adapter
16493L
Output
Low
Description
Keysight B1505A Configuration and Connection Guide
N1259A
HVSMU2
HCSMU3 F
Module
Selecto
HCSMU3 S
HCSMU2 F
HCSMU2 S
GNDU2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents