Connection Guide for Wafer Prober and Your Own Test Fixture

To Connect Ultra High Voltage Expander

To Connect Ultra High Voltage Expander
The N1268A ultra high voltage expander is used to configure the ultra high voltage unit
(UHVU) with the MCSMU modules. And the UHVU can be connected to prober as shown
in
near the B1505A. And the N1269A should be installed near prober as shown in
"Connection
Figure 5-19
To Configure UHVU
WARNING
Connect a wire from an electrical ground (safety ground) to the earth terminal of the
N1268A.
Table 5-30
To Connect N1268A/N1269A Input
B1505A terminals
MCSMU
Force
16494A-001 Triaxial cable, 1.5 m
Sense
16494A-001 Triaxial cable, 1.5 m
1
Force
16494A-001 Triaxial cable, 1.5 m
MCSMU
Sense
16494A-001 Triaxial cable, 1.5 m
16493J-001 Interlock cable, 1.5 m
Interlock
Digital I/O
16493G-001 Digital I/O connection cable, 1.5 m
Power cable, for connecting to power line
Wire, 1 ea., for earthing
1. HCSMU is substitutable. Then use 16493S HCSMU cable, N1254A-103 adapter for Force, and
N1254A-517 adapter for Sense.
5-38
Figure
5-19, by using the N1269A connection adapter. The N1268A should be installed
Overview".
B1505A
N1268A UHV expander
MCSMU
I-control
MC/HCSMU
V-control
Interlock
Interlock Input
Digital I/O
Digital I/O
MCSMU
MCSMU
N1269A UHV connection adapter
Required cables
Interlock circuit
Interlock Output
High
Low
Keysight B1505A Configuration and Connection Guide
Drain
DUT
Chuck
Gate
Source
N1268A input
Force
I Control
Sense
Force
V Control
Sense
Interlock Input
Digital I/O Input
LINE
(earth)

Hide quick links: