To Install an Interlock Circuit

The interlock circuit is designed to prevent electrical shock when a user touches the
measurement terminals.
You must install an interlock circuit on a shielding box to prevent hazardous voltages when
the door of the shielding box is open.
Figure 5-24
on your DUT interface. And
circuit.
WARNING
Potentially hazardous voltages may be present at the Force, Guard, and Sense
terminals when the interlock terminals are shorted.
Figure 5-24
Interlock Connector Pin Assignments
Table 5-38
Recommended Parts
Keysight B1505A Configuration and Connection Guide
shows the pin assignments of the interlock connector that should be mounted
6
1
5
2
4
3
Connector side
Keysight Part
No.
N1254A-402
Interlock micro switch
1252-1419
Interlock connector (6 pin, female)
1450-0641
LED (V
Wire 24AWG, 600 V, 150 degree C or equivalent
Connection Guide for Wafer Prober and Your Own Test Fixture
Table 5-38
lists the required parts to make the interlock
Interlock switch
1
2
3
Wiring side
Description
2.1 V @ I
= 10 mA)
F
F
To Install an Interlock Circuit
6
5
LED
4
Quantity
2
1
1
5-47