N1265A Connection Guide
Output Connection
3. Set the DUT on the socket.
4. Close the fixture cover and perform measurement.

Universal Socket Module (N1265A-011)

This is a blank module, kind of a do-it-yourself kit for supporting variety of packaged
devices. This module can be used by mounting your desired socket or packaged device and
making connections same as the N1265A-010 Inline Package Socket Module.
To use this module, see the following simple instruction. For the component locations and
dimensions, see
page 3-17
Figure 3-8
Universal Socket Module
82 mm
(Nut pitch: 69 mm)
91 mm (Nut pitch: 78 mm)
1. Prepare the following parts.
2. Cut the blank board in 90 mm  81 mm square.
3. Make four screw holes on the board. The holes should be 6 mm inside from the edge.
4. Fix the board to the blank module.
5. Remove the cover bottom of the blank module.
3-18
HC/MC/DHCSMU via SMU 3 terminals should be connected as follows.
SMU 3 High Force and Sense to DUT high
SMU 3 Low Force and Sense to DUT low
GNDU Force and Sense to DUT low
The GNDU signals will appear at the Selector Output Low terminals as shown in
Table
3-7.
Figure
3-8. Also see
to perform measurement.
N1265A Opt 011
Emi er / Source
Base / Gate
Collector / Drain
Force
Sense
Force
Sense
1
Force
Sense
2
2
1
3
4
5
6
7
3
4
8
Universal Socket Module
blank board suitable for mounting the socket or packaged device
screw M3, 4 ea., for fixing the blank board on the socket module
wire, adequate length and quantity, for making connections
hexlobe (torx type) screwdriver T-10
socket, if you use, and packaged device under test (DUT)
"Inline Package Socket module (N1265A-010)" on
165 mm
5
6
129 mm