N1259A Connection Guide
Output Connection
To Use High Voltage Bias-T
The high voltage bias-T is used to perform capacitance or impedance measurements using
DC bias of up to ± 3000 V. The bias-T High terminal is internally connected to the MF
CMU Hpot and Hcur input connectors. And the bias-T Low terminal is internally
connected to the MF CMU Lpot and Lcur input connectors.
Figure 2-9
High Voltage Bias-T Connection Example
2-18
Required parts:
N1254A-512 SHV(plug)-SHV(plug) cable, 2 ea.
N1254A-513 SHV(jack)-banana adapter, 2 ea.
N1254A-508 or N1254A-509 connection wire, 2 ea.
Connection:
See
Figure
2-9.
1. Attach the SHV-banana adapter to the SHV cable, and make two cables.
2. Connect a cable (solid line) between the bias-T Low terminal and the low terminal
of DUT.
3. Connect a cable (solid line) between the bias-T High terminal and the high terminal
of DUT.
4. Connect a wire (dotted line) between a SMU Force output terminal and the DC Bias
Input High terminal. Use MPSMU for DC bias of up to ± 100 V, HPSMU for ± 200
V, or HVSMU for ± 3000 V.
5. Connect a wire (dotted line) between the SMU Guard output terminal and the DC
Bias Input Guard terminal.
N1259A
Opt 022
N1259A
1
2
100 kΩ
±3 kV
Max
Max: 6.4 W
Force
Guard
1
Sense
Force
Force
Sense
Sense
Guard
±40 V
Max
±3 kV
Max
Low
High
Module Selector Output
Force
Guard
Opt 010
Kelvin Connection
Force
Sense
Force
Sense
1 2 3
±3 kV
Max
Inline Package Socket (3 Pin)
Keysight B1505A Configuration and Connection Guide
N1259 A
Opt 021
1
2
1 MΩ
±3 kV
Max
Max: 9 W
2
DC Bias Input
±3 kV
Max
Guard
Force
High
3
±3 kV
Max
Low
±25 V
Max
MF CMU
High Voltage Bias –Tee