Keysight Technologies B1505A Configuration Manual page 155

Power device analyzer/curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Table 5-3
To Connect N1258A Module Selector
N1266A
*5
B 1505A
16494A/16493K
16493S
16493T
16493L
16493G
*5: 16494A, 4 ea., 16493L, 16493T, and 16493G
to configure HVMCU.
B1505A terminals
Force
16494A Triaxial cable, 3 m or 1.5 m, 1 ea., for the non-Kelvin
HPSMU
connection, connect it between the Force connectors.
or
Sense
MPSMU
For the Kelvin connection, prepare additional one more cable and
connect it between the Sense connectors. Or use 16493K Kelvin
triaxial cable instead of 16494A.
Use 16493K cable if the High output terminals are connected to
the prober chuck.
Force
16493S HCSMU cable, 3 m or 1.5 m, 1 ea. for HCSMU
HCSMU
or
16493S HCSMU cable, 3 m or 1.5 m, 2 ea. and 16493S-021 Dual
Sense
DHCSMU
HCSMU adapter, 1 ea. for DHCSMU
Force
16493T HVSMU cable, 3 m or 1.5 m, 1 ea.
HVSMU
HVMCU can be used instead of HVSMU. See
16493L GNDU cable, 3 m or 1.5 m, 1 ea. Also see
GNDU
Digital I/O
16493G Digital I/O connection cable, 3 m or 1.5 m, 1 ea.
N1258A terminals
Force
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
Low
Sense
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
Force
Coaxial cable with HV(plug) connector. N1254A-506 can be used.
High
Use coaxial cable to minimize affect of inductance.
Sense
Triaxial cable with HV(plug) connector. N1254A-505 can be used.
LINE
Power cable, 1 ea., for connecting to power line
1. Non-Kelvin connection is not allowed. GNDU Force and Sense are internally connected to Low Sense.
See
Table 5-22 on page
Keysight B1505A Configuration and Connection Guide
*1: 1252-1419
*2: N1254A-506 or other cable, 1 ea.
*3: N1254A-505 or other cable, 1 ea.
*4: N1254A-503 or other cable, 2 ea.
16493J
*1
Interlock circuit
N1258A
Required cables
Required cables/Description
5-31.
Connection Guide for Wafer Prober and Your Own Test Fixture
Force *2
Sense
*3
High
Low
Table
5-2.
Table
5-2.
Connection Overview
Shielding box
DUT interface
*4
Force
Sense
N1258A terminals
Force
HPSMU
Sense
Force
HCSMU
Sense
HVSMU
GNDU
Digital I/O
Connect to
DUT low terminal
1
(ex: source)
DUT high terminal
1
(ex: drain)
site AC power
5-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents